Model No.︰AV-2000B1
Brand Name︰HIGHSUN
Country of Origin︰China
Unit Price︰US $ 1900 / pc
Minimum Order︰1 pc
AV-2000B1
The AV-2000B1 multi-purpose life-support machines are suitable for the children, the adults, widely used in the
hospital wards, ICU, A & E.
Main technical parameters
Main technical parameters
Ventilator
■Display mode 5.7-inch 8.4-inch high clear display
■Ventilation mode
IPPV(intermittent positive pressure ventilation)
SIPPV(synchronized intermittent positive pressure ventilation) SIMV(synchronized intermittent mandatory ventilation)
PEEP(expiration end positive pressure)
MANUAL(manual work)
SIGH
■Ventilation parameters
Tidal volume 50~1500ml
Frequency 4~60/min
Inspiratory time/expiratory time ratio 2:1~1:8
Trigger sensitivity -0.1~-1.0KPa
PEEP 0~1.0KPa
Pressure control 1~6KPa
Sigh 1.5 inspiratory times
■Ventilation monitoring
Tidal volume、Ventilation capacity、IPPV frequency、Breathing frequency、Inspiratory ime/expiratory time ratio、Peak &#118alue of airway pressure、Pressure-time wave、Flow rate-time wave、Breathing frequency、End-expiratory positive pressure、 Inspiratory trigger pressure、 (Inspiratory trigger
pressure、Inspiratory pause)
■(Oxygen concentration monitor 21%~100% )
■Safety alarm systems
(Oxygen concentration alarm ) Upper limit setting range 50%~1 Low limit setting range 15%~50%
Airway pressure alarm Upper limit setting range 1.5~6.0KPa
Low limit setting range 0.2~3.0KPa
Minute volume alarm Upper limit setting range 2.0~30L/min
Low limit setting range 1~12L/min
Asphyxia alarm The machine will give sound and light alarm if there is no
tidal volume more than 15s
Power failure alarm
■Power supply AC 220V±10% 50Hz±1Hz
UPS、 Battery (choose and match)
Attention: In the above parenthesis the content is 8.4 inch colored tablet life-support machine’s parameter
Payment Terms︰ TT/LC